mpMINIPURE BASIC
OTO / UP OTO
General
Applicable to laboratory, analyzers , biochemistry , hormon-elisa tests , reseach , QC , , high performance analysis.
BASIC OTO produces CAP , NCCLS , ASTM ve ISO 3696 (TYPE 2 ve TYPE 3) BASIC UP OTO produces ASTM TYPE I , NCCLS TYPE I ve ISO 3696 TYPE I quality water up to 50 liters per day from tap water directly.

UYGULAMA ALANLARI - Basic OTO
- Replacement for distillation units
- General lab analysis
- Autoanalyzers
- AAS
- Final rinse
- Autoclaves
- Test chambers
- Buffer preperation
- Culture media
- Photometer
- Spectrophotometer
- Sitology
- Pathology
- Histology
- Elektrophoresis
Applications - with Basic UP OTO
- HPLC
- IC
- GC/MS
- TOC analysis
- DNA analysis
- IVF
- 2D elektrophoresis
- Critical Cell cultures
- Pyrogen sensitive applications
MPMINIPURE BASIC® OTO ve UP OTO
- Flushable mechanical filter reduces cartridge consumption
- Automatic external softener reduces hardness of tap water and prolongs life of RO membranes
- High capacity combined activated carbon + 5 micron cartridge holds particles , organic matter and free chlorine
- Qualified RO membranes rejects 96% of ions and 100% of particles,organic matter and bacteria in softened and filtered water Dirty water is sent to drain
- Purified water is collected in internal 20 liter capacity PP pure water tank
- in BASIC UP OTO models purified water passes trough Ultraviyole ünit, in which bacteria and organic matter is destroyed
- Special magnetic level sensors stops production when tank is filled with purified water, Low level sensor also protects pure water pump , if there is no water in the tank
- Purified water in the tank is delivered with adjustable pressure to mixed bed resin tanks
- First tank , filled with mixed resin holds ions in the purified water , Second tank filled with NRW type resin (additionally TOC reducing resin in UP models) holds all remaining anions and cations and provides ultrapure water at the outlet on the front panel and the back
- User can obtain ultrapure water in speed due to pure water pump
- Optional 0,2 micron filters placed on the outlet valve holds bacteria and endotoxins
- Each MPMINIPURE is equipped with very reliable MES RO/DI CONTROLLER 08
- Microprosesor electronics controls internal tank , stops the production when full and activates again when Water level is reduced, Flushes the membranes
- Alarms when there is no water in the tank and protects pure water pump
- Activates the pure water pump ,if the water level is suitable
|
*Different capacity pure water pumps are available as options |
|||||||||||||||||||||||||||||||||||||||||||||||||||
Specifications are subject to change without notice under ligth of continous improvements. |
||||||||||||||||||||||||||||||||||||||||||||||||||||